Chapter B: SETTINGS 4 TECHNICAL SETTINGS
Technical manual Planmeca ProOne 73
Enable programs (T3720)
Here you can select or deselect the following programs or features:
Enable Autofocus
To activate this feature touch the
Enable Autofocus
field. The field will be highlighted.
To disable this feature touch the field again. The field will turn white and it will then not
be possible to take any Autofocus exposures.
Enable TMJ double PA
This program is activated (highlighted) by default. To disable this program touch the
Enable TMJ double PA
field. The field will turn white and it will then not be possible to
take any TMJ double PA exposures.
Enable TMJ double lateral-PA
By default, the TMJ double lateral-PA program is part of the “Advanced programs”
package, and the
Enable TMJ double lateral-PA
field is highlighted when the
“Advanced programs” package has been activated.
Note. The option will be shown on the
Enable programs (T3720)
display only if the
“Advanced programs” package has been activated on the
Program licences (P2100)
display.
To disable the TMJ double lateral-PA program, touch the
Enable TMJ double lateral-
PA
field. The field will turn white and it will then not be possible to take any TMJ
double lateral-PA exposures. All other exposure programs in the “Advanced
programs” package will still be available.
Enable Cross sections
By default, cross-sectional programs are part of the “Advanced programs” package,
and the
Enable Cross sections
field is highlighted when the “Advanced programs”
package has been activated.
Note. The option will be shown on the
Enable programs (T3720)
display only if the
“Advanced programs” package has been activated on the
Program licences (P2100)
display.
To disable cross-sectional programs, touch the
Enable Cross sections
field. The field
will turn white and it will then not be possible to take any cross-sectional exposures.
All other exposure programs in the “Advanced programs” package will still be
available.
Accept your selection by touching the green check mark button in the bottom right corner
of the display. To exit the display without saving the changes, touch the red cross button in
the bottom left corner of the display.

Table of Contents

Other manuals for Planmeca ProOne

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProOne and is the answer not in the manual?

Planmeca ProOne Specifications

General IconGeneral
BrandPlanmeca
ModelProOne
CategoryMedical Equipment
LanguageEnglish

Related product manuals