Remote Control
R&S
®
FSPN
331User Manual 1179.4363.02 ─ 01
Example: //Query residual AM
FETC:RANG2:PNO:RAM?
would return, e.g.
0.01196555513888597
Usage: Query only
Manual operation: See "Integrated Measurements" on page 89
FETCh<n>[:RANGe<j>]:PNOise<t>:RFM?
This command queries the residual FM.
Suffix:
<n>
.
Window
<j> Integration range
<t> Trace
Return values:
<Frequency> <numeric value>
Default unit: Hz
Example: //Query residual FM
FETC:RANG2:PNO:RFM?
would return, e.g.
88110000
Usage: Query only
Manual operation: See "Integrated Measurements" on page 89
FETCh<n>[:RANGe<j>]:PNOise<t>:RMS?
This command queries the residual RMS jitter.
Suffix:
<n>
.
Window
<j> Integration range
<t> Trace
Return values:
<Jitter> <numeric value>
Default unit: s
Example: //Query RMS jitter
FETC:RANG2:PNO2:RMS?
would return, e.g.
0.02175
Usage: Query only
Manual operation: See "Integrated Measurements" on page 89
Measurements and Result Displays