Remote Control
R&S
®
FSPN
382User Manual 1179.4363.02 ─ 01
Example: See CONFigure:VCO:SWEep:SOURce.
Manual operation: See "Defining the sweep range" on page 128
CONFigure:VCO:SWEep:STOP <Value>
This command defines the stop value of the tuning range.
Parameters:
<Value> <numeric value>
The unit depends on whether you tune the VCO with a voltage
or current (see CONFigure:VCO:SWEep:SOURce).
*RST: 0
Default unit: V or A
Example: See CONFigure:VCO:SWEep:SOURce.
Manual operation: See "Defining the sweep range" on page 128
11.6.8
Spot Noise vs Tune Configuration
● Sweep Configuration.............................................................................................382
● Trace Configuration...............................................................................................384
11.6.8.1 Sweep Configuration
CONFigure:SNTune:SWEep:DELay:INITial.......................................................................382
CONFigure:SNTune:SWEep:DELay[:POINt]..................................................................... 383
CONFigure:SNTune:SWEep:POINts................................................................................383
CONFigure:SNTune:SWEep:SOURce..............................................................................383
CONFigure:SNTune:SWEep:STARt................................................................................. 384
CONFigure:SNTune:SWEep:STOP..................................................................................384
CONFigure:SNTune:SWEep:DELay:INITial <SettlingTime>
This command defines an initial delay time that allows the DUT to settle before the first
measurement starts.
Parameters:
<SettlingTime> <numeric value>
*RST: 10 ms
Default unit: s
Example: //Define an initial settling time of a 100 ms.
CONF:VCO:SWE:DEL:INIT 100MS
Manual operation: See "Defining the settling time" on page 129
Common Measurement Settings