EasyManua.ls Logo

R&S FSPN - Common Measurement Settings; Remote Commands to Configure the Input Source

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
337User Manual 1179.4363.02 ─ 01
// Turn on marker 1 in window 1
CALC1:MARK1:STAT ON
// Set marker 1 stimulus
CALC1:MARK1:X 300 kHz
// Query marker 1 result
CALC1:MARK1:Y?
11.6 Common Measurement Settings
● Remote Commands to Configure the Input Source.............................................. 337
● Output................................................................................................................... 339
● Remote Commands to Configure User Ports........................................................348
● Remote Commands to Configure Level Characteristics....................................... 350
● Remote Commands to Configure the Frequency..................................................356
● Phase Noise Measurement Configuration............................................................ 360
● VCO Characterization Configuration.....................................................................379
● Spot Noise vs Tune Configuration........................................................................ 382
● Transient Analysis.................................................................................................386
11.6.1 Remote Commands to Configure the Input Source
● RF Input................................................................................................................ 337
11.6.1.1 RF Input
INPut<ip>:COUPling.......................................................................................................337
INPut<ip>:SELect...........................................................................................................338
INPut<1|2>:LOSCillator:SOURce.....................................................................................338
INPut<1|2>:LOSCillator:SOURce:EXTernal:LEVel............................................................. 338
INPut<ip>:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType> AC | DC
AC
AC coupling
DC
DC coupling
*RST: AC
Example:
INP:COUP DC
Common Measurement Settings