R&S FSPN format
Format is also compatible to the R&S FSL and R&S FSV family
*RST: not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation: See "*IDN Format" on page 284
SYSTem:PRESet:FILTer <FilterType>
This command selects the resolution filter type that is selected after a preset in the
Spectrum application.
Parameters:
<FilterType> NORMal
Selects 3 dB filter.
NOISe
Selects 3 dB filter.
(NORMal and NOISe have the same effect.)
PULSe
Selects 6 dB filter.
*RST: NORMal
Example: //Select the 6 dB filters as the default filter type
SYST:PRES:FILT PULS
11.9.8 Using Service Functions
DIAGnostic:SERVice:SFUNction......................................................................................507
DIAGnostic:SERVice:SFUNction:LASTresult?................................................................... 508
DIAGnostic:SERVice:SFUNction:RESults:DELete............................................................. 508
DIAGnostic:SERVice:SFUNction:RESults:SAVE................................................................508
DIAGnostic:SERVice:SINFo?.......................................................................................... 508
SYSTem:PASSword:RESet............................................................................................. 509
SYSTem:PASSword[:CENable]........................................................................................509
DIAGnostic:SERVice:SFUNction <ServiceFunction>
This command starts a service function.
The service functions are available after you have entered the level 1 or level 2 system
password.
Parameters for setting and query:
<ServiceFunction> String containing the ID of the service function.
The ID of the service function is made up out of five numbers,
separated by a point.
Configuring the R&S FSPN