477User Manual 1179.4363.02 ─ 01
11.9 Configuring the R&S FSPN
● Basic Instrument Setup.........................................................................................477
● Configuring the Reference Frequency.................................................................. 478
● Calibrating the R&S FSPN and Checking Temperature........................................483
● Customizing the Screen Layout............................................................................ 486
● Configuring Network and Remote Control............................................................ 492
● Configuring HUMS................................................................................................ 497
● Checking the System Configuration......................................................................503
● Using Service Functions....................................................................................... 507
● Reference: Command Set of 5052 Emulation.......................................................509
● Reference: Command Set of Emulated PSA Models............................................511
11.9.1 Basic Instrument Setup
SYSTem:CLOGging....................................................................................................... 477
SYSTem:REBoot............................................................................................................477
SYSTem:SHUTdown...................................................................................................... 477
SYSTem:DATE.............................................................................................................. 478
SYSTem:TIME...............................................................................................................478
SYSTem:CLOGging <State>
This command turns logging of remote commands on and off.
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
Writes all remote commands that have been sent to a file.
The destination is C:\R_S\INSTR\ScpiLogging\
ScpiLog.<no.>.
where <no.> is a sequential number
A new log file is started each time logging was stopped and is
restarted.
OFF | 0
*RST: 0
Manual operation: See "I/O Logging" on page 284
SYSTem:REBoot
This command reboots the instrument, including the operating system.
SYSTem:SHUTdown
This command shuts down the instrument.