386User Manual 1179.4363.02 ─ 01
Example: //Remove spurs
SPUR:SUPP ON
Manual operation: See "Hiding spurs" on page 120
[SENSe:]SPURs:THReshold <Threshold>
This command defines a level threshold for spur removal.
Prerequisites for this command
Select Spot Noise vs Tune measurement (CONFigure:SNTune:MEASurement[:
STATe]).
Turn on spur removal (DISPlay[:WINDow<n>]:TRACe<t>:SPURs:THReshold
).
The threshold applies to all traces. If you need to define a threshold for each trace sep-
arately (possible in all measurements except Spot Noise vs Tune), use DISPlay[:
WINDow<n>]:TRACe<t>:SPURs:THReshold.
Parameters:
<Threshold> <numeric value>
*RST: 0
Default unit: dB
Example: //Remove spurs
SPUR:SUPP ON
SPUR:THR 5
Manual operation: See "Hiding spurs" on page 120
11.6.9 Transient Analysis
Frequency Configuration.......................................................................................386
Amplitude Configuration........................................................................................389
Data Acquisition.................................................................................................... 390
Trigger Configuration............................................................................................ 391
Y-Axis Scale.......................................................................................................... 394
Settling Time......................................................................................................... 398
Frequency Deviation............................................................................................. 399
Trace Configuration...............................................................................................401
Marker Configuration.............................................................................................402
Limit Lines.............................................................................................................402
11.6.9.1 Frequency Configuration
Commands to configure the frequency described elsewhere.
Center frequency: [SENSe:]FREQuency:CENTer