Remote Control
R&S
®
FSPN
385User Manual 1179.4363.02 ─ 01
Prerequisites for this command
●
Select Spot Noise vs Tune measurement (CONFigure:SNTune:MEASurement[:
STATe]).
●
Turn on trace smoothing ([SENSe:]SMOothing[:STATe]).
This command smoothes all traces. If you need to smooth each trace separately (pos-
sible in all measurements except Spot Noise vs Tune), use DISPlay[:WINDow<n>]:
TRACe<t>:SMOothing:APERture.
Parameters:
<Value> <numeric value>
Range: 1 to 20
*RST: 1
Default unit: PCT
Example: //Smooth the traces by a magnitude of 10 %
SMO ON
SMO:APER 10
Manual operation: See "Smoothing traces" on page 157
[SENSe:]SMOothing[:STATe] <State>
This command turns trace smoothing for a specific trace on and off.
When you turn on trace smoothing, you can define the smoothing magnitude
with[SENSe:]SMOothing:APERture .
Prerequisites for this command
●
Select Spot Noise vs Tune measurement (CONFigure:SNTune:MEASurement[:
STATe]).
This command smoothes all traces. If you need to smooth each trace separately (pos-
sible in all measurements except Spot Noise vs Tune), use DISPlay[:WINDow<n>]:
TRACe<t>:SMOothing[:STATe].
Parameters:
<State> ON | OFF | 1 | 0
Example: //Smooth the traces by a magnitude of 10 %
SMO ON
SMO:APER 10
Manual operation: See "Smoothing traces" on page 157
[SENSe:]SPURs:SUPPress <State>
This command turns spur removal for all traces and windows on and off.
Parameters:
<State> ON | OFF | 1 | 0
Common Measurement Settings