Remote Control
R&S
®
FSPN
459User Manual 1179.4363.02 ─ 01
Manual operation: See " Items: " on page 192
MMEMory:SELect:CHANnel[:ITEM]:WEIGhting <State>
MMEMory:SELect[:ITEM]:WEIGhting <State>
Parameters:
<State>
11.8.3 Saving and Loading Instrument Settings
MMEMory:CLEar:ALL.....................................................................................................459
MMEMory:CLEar:STATe................................................................................................. 459
MMEMory:LOAD:AUTO..................................................................................................459
MMEMory:LOAD:STATe..................................................................................................460
MMEMory:STORe<1|2>:STATe....................................................................................... 461
MMEMory:STORe<1|2>:STATe:NEXT..............................................................................461
MMEMory:STORe<1|2>:TYPE........................................................................................ 461
SYSTem:PRESet........................................................................................................... 462
SYSTem:PRESet:CHANnel[:EXEC]................................................................................. 462
MMEMory:CLEar:ALL
This command deletes all instrument configuration files in the current directory.
You can select the directory with MMEMory:CDIRectory on page 452.
Example:
MMEM:CLE:ALL
MMEMory:CLEar:STATe <1>, <FileName>
This command deletes an instrument configuration file.
Parameters:
<1>
<FileName> String containing the path and name of the file to delete.
The string may or may not contain the file's extension.
Example:
MMEM:CLE:STAT 1,'TEST'
MMEMory:LOAD:AUTO <1>, <FileName>
This command restores an instrument configuration and defines that configuration as
the default state.
The default state is restored after a preset (*RST) or after you turn on the R&S FSPN.
Parameters:
<1>
<FileName> 'Factory'
Restores the factory settings as the default state.