EasyManua.ls Logo

R&S FSPN - Configuring the Reference Frequency

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
478User Manual 1179.4363.02 ─ 01
SYSTem:DATE <Year>, <Month>, <Day>
Configures the date on the instrument.
Parameters:
<Year>
<Month>
<Day>
Example:
SYST:DATE 2020,04,23
Manual operation: See " Set Date and Time " on page 217
SYSTem:TIME
<Year>, <Month>, <Day>
Configures the time on the internal real-time clock on the instrument.
Parameters:
<Hour> Range: 0 to 23
<Minutes> Range: 0 to 59
<Seconds> Range: 0 to 59
Example:
SYST:TIME 10,52,33
Manual operation: See " Set Date and Time " on page 217
11.9.2 Configuring the Reference Frequency
[SENSe:]ROSCillator:COUPling:BANDwidth..................................................................... 478
[SENSe:]ROSCillator:COUPling:BANDwidth:MODE.......................................................... 479
[SENSe:]ROSCillator:COUPling:MODE............................................................................479
[SENSe:]ROSCillator:LBWidth.........................................................................................479
[SENSe:]ROSCillator:O640.............................................................................................480
SOURce<si>:EXTernal<ext>:ROSCillator:EXTernal:FREQuency.........................................480
[SENSe:]ROSCillator:OSYNc.......................................................................................... 480
[SENSe:]ROSCillator:SOURce........................................................................................ 481
[SENSe:]ROSCillator:SOURce:EAUTo?........................................................................... 481
[SENSe:]ROSCillator:TRANge.........................................................................................482
[SENSe:]ROSCillator:EXTernal:FALLback........................................................................ 482
[SENSe:]ROSCillator:COUPling:BANDwidth <Bandwidth>
This command defines coupling bandwidth for the internal reference frequency.
Prerequisites for this command
Options R&S FSPN-B60 or -B61 must be available.
Select manual bandwidth mode ([SENSe:]ROSCillator:COUPling:
BANDwidth:MODE)
Configuring the R&S
FSPN

Table of Contents

Related product manuals