Remote Control
R&S
®
FSPN
486User Manual 1179.4363.02 ─ 01
SOURce<si>:TEMPerature:FRONtend
This command queries the current frontend temperature of the R&S FSPN.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the STATUs:QUEStionable:TEMPerature register indicates a
possible deviation.
Suffix:
<si>
.
irrelevant
Return values:
<Temperature> Temperature in degrees Celsius.
Example:
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
11.9.4 Customizing the Screen Layout
● Configuring Screen Elements............................................................................... 486
● Selecting Colors and Schemes.............................................................................489
● CMAP Suffix Assignment......................................................................................491
11.9.4.1 Configuring Screen Elements
Useful commands to configure screen elements described elsewhere:
●
DISPlay[:WINDow<n>]:MTABle
●
DISPlay:FORMat
DISPlay:ANNotation:CBAR............................................................................................. 486
DISPlay:ANNotation:FREQuency.....................................................................................487
DISPlay:SBAR[:STATe]...................................................................................................487
DISPlay:SKEYs[:STATe]................................................................................................. 487
DISPlay:TBAR[:STATe]...................................................................................................487
DISPlay:TOUChscreen[:STATe].......................................................................................488
DISPlay[:WINDow<n>]:TIME...........................................................................................488
DISPlay[:WINDow<n>]:TIME:FORMat..............................................................................488
SYSTem:DISPlay:FPANel[:STATe]................................................................................... 489
DISPlay:ANNotation:CBAR <State>
This command hides or displays the channel bar information.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Configuring the R&S FSPN