Remote Control
R&S
®
FSPN
488User Manual 1179.4363.02 ─ 01
DISPlay:TOUChscreen[:STATe] <State>
This command controls the touch screen functionality.
Parameters:
<State> ON | FRAMe | OFF
ON | 1
Touch screen is active for entire screen
OFF | 0
Touch screen is inactivate for entire screen
FRAMe
Touch screen is inactivate for the diagram area of the screen,
but active for softkeys, toolbars and menus.
*RST: 1
Example:
DISP:TOUC:STAT ON
Manual operation: See " Deactivating and Activating the Touchscreen "
on page 217
DISPlay[:WINDow<n>]:TIME <State>
This command adds or removes the date and time from the display.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
DISP:TIME ON
Manual operation: See " Date and Time " on page 219
DISPlay[:WINDow<n>]:TIME:FORMat <Format>
This command selects the time and date format.
Suffix:
<n>
.
irrelevant
Parameters:
<Format> US | DE
DE
dd.mm.yyyy hh:mm:ss
24 hour format.
US
mm/dd/yyyy hh:mm:ss
12 hour format.
*RST: DE
Configuring the R&S FSPN

Table of Contents