11.6.5 Remote Commands to Configure the Frequency
[SENSe:]ADJust:CONFigure:FREQuency:AUTosearch[:STATe].......................................... 356
[SENSe:]ADJust:CONFigure:FREQuency:COUNt............................................................. 356
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:HIGH........................................................357
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:LOW........................................................ 357
[SENSe:]FILTer:LPASs:FREQuency:MANual.................................................................... 357
[SENSe:]FILTer:LPASs[:STATe]....................................................................................... 358
[SENSe:]FREQuency:CENTer......................................................................................... 358
[SENSe:]FREQuency:STARt........................................................................................... 359
[SENSe:]FREQuency:STOP............................................................................................359
[SENSe:]POWer:RLEVel:MODE......................................................................................360
[SENSe:]SWEep:CAPTure:RANGe..................................................................................360
[SENSe:]ADJust:CONFigure:FREQuency:AUTosearch[:STATe] <State>
This command turns the automatic (carrier) frequency search on and off.
When you turn of the automatic frequency search, you can define the signal frequency
with [SENSe:]FREQuency:CENTer.
Parameters:
<State> ON | OFF | 1 | 0
Example: //Define a signal frequency of 100 MHz
ADJ:CONF:FREQ:AUT OFF
FREQ:CENT 100MHZ
Manual operation: See "Searching for a signal" on page 107
[SENSe:]ADJust:CONFigure:FREQuency:COUNt <State>
This command turns verification of the signal frequency on and off.
Prerequisites for this command
Turn off automatic signal search ([SENSe:]ADJust:CONFigure:FREQuency:
AUTosearch[:STATe]).
Parameters:
<State> OFF | 0
Turns off the frequency counter and measures the frequency
you have entered without verifiying it.
ON | 1
Turns on the frequency counter and verfifies the frequency you
have entered.
*RST: OFF (automatic signal search is on)
Example: //Turn off the frequency counter
ADJ:CONF:FREQ:AUT OFF
ADJ:CONF:FREQ:COUN OFF
FREQ:CENT 93.423MHZ
Manual operation: See "Searching for a signal" on page 107