Remote Control
R&S
®
FSPN
355User Manual 1179.4363.02 ─ 01
[SENSe:]ADJust:CONFigure:LEVel:THReshold <Level>
This command defines the threshold of the signal search.
Prerequisites for this command
●
Turn on the automatic signal search with [SENSe:]ADJust:CONFigure:
FREQuency:AUTosearch[:STATe]
Parameters:
<Level> <numeric value>
Default unit: dBm
Example: //Define a threshold of -10 dBm
ADJ:CONF:LEV:THR -10
Manual operation: See "Searching for the signal level" on page 103
[SENSe:]POWer:RLEVel?
This command queries the currently measured signal level.
Return values:
<Level> <numeric value>
Default unit: dBm
Example: //Query signal level
POW:RLEV?
would return, e.g.
-9.28192
Usage: Query only
Manual operation: See "Searching for the signal level" on page 103
[SENSe:]POWer:RLEVel:OFFSet <Offset>
This command defines an arithmetic level offset. The offset is applied to the signal
level.
The level offset takes external attenuation or gain into account.
Parameters:
<Offset> <numeric value>
Range: -100 to 200
*RST: 0
Default unit: dB
Example: //Define a level offset of 20 dB
POW:RLEV:OFFS 20
Manual operation: See "Shifting the level" on page 103
Common Measurement Settings