Remote Control
R&S
®
FSPN
357User Manual 1179.4363.02 ─ 01
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:HIGH <Frequency>
This command defines the upper limit of the frequency search range.
Prerequisites for this command
●
Turn on the automatic signal search with [SENSe:]ADJust:CONFigure:
FREQuency:AUTosearch[:STATe]
Parameters:
<Frequency> <numeric value>
Range: See data sheet
*RST: 8 GHz
Default unit: Hz
Example: //Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:HIGH 500MHZ
Manual operation: See "Searching for a signal" on page 107
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:LOW <Frequency>
This command defines the lower limit of the frequency search range.
Prerequisites for this command
●
Turn on the automatic signal search with [SENSe:]ADJust:CONFigure:
FREQuency:AUTosearch[:STATe]
Parameters:
<Frequency> <numeric value>
Range: See data sheet
*RST: 1 MHz
Default unit: Hz
Example: //Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:UPP 500MHZ
Manual operation: See "Searching for a signal" on page 107
[SENSe:]FILTer:LPASs:FREQuency:MANual <Frequency>
This command defines the cutoff frequency of the low pass filter you can use to mea-
sure small carrier frequencies.
Prerequisites for this command
●
Turn on low pass filter ([SENSe:]FILTer:LPASs[:STATe]).
Parameters:
<Frequency> <numeric value>
*RST: 1 MHz
Default unit: Hz
Common Measurement Settings