EasyManua.ls Logo

R&S FSPN - Page 357

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
357User Manual 1179.4363.02 ─ 01
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:HIGH <Frequency>
This command defines the upper limit of the frequency search range.
Prerequisites for this command
Turn on the automatic signal search with [SENSe:]ADJust:CONFigure:
FREQuency:AUTosearch[:STATe]
Parameters:
<Frequency> <numeric value>
Range: See data sheet
*RST: 8 GHz
Default unit: Hz
Example: //Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:HIGH 500MHZ
Manual operation: See "Searching for a signal" on page 107
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:LOW <Frequency>
This command defines the lower limit of the frequency search range.
Prerequisites for this command
Turn on the automatic signal search with [SENSe:]ADJust:CONFigure:
FREQuency:AUTosearch[:STATe]
Parameters:
<Frequency> <numeric value>
Range: See data sheet
*RST: 1 MHz
Default unit: Hz
Example: //Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:UPP 500MHZ
Manual operation: See "Searching for a signal" on page 107
[SENSe:]FILTer:LPASs:FREQuency:MANual <Frequency>
This command defines the cutoff frequency of the low pass filter you can use to mea-
sure small carrier frequencies.
Prerequisites for this command
Turn on low pass filter ([SENSe:]FILTer:LPASs[:STATe]).
Parameters:
<Frequency> <numeric value>
*RST: 1 MHz
Default unit: Hz
Common Measurement Settings