Remote Control
R&S
®
FSPN
394User Manual 1179.4363.02 ─ 01
11.6.9.5 Y-Axis Scale
Commands to scale the y-axis in the frequency result display described elsewhere.
Center frequency: [SENSe:]FREQuency:CENTer
Start frequency: [SENSe:]FREQuency:STARt
Stop frequency: [SENSe:]FREQuency:STOP
Span: [SENSe:]FREQuency:SPAN
Stepsize: [SENSe:]FREQuency:CENTer:STEP:LINK
Stepsize factor: [SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
Commands to scale the y-axis in the phase result display described elsewhere.
Mode: DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO
Auto scale once: DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling..........................................................394
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint:MODE....................................394
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint[:X]..........................................395
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE.........................................................395
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision.................................................... 396
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition................................................... 396
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue.......................................................397
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:UNIT...........................................................397
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling <Mode>
This command selects the AF coupling mode.
Suffix:
<n>
.
1..n
Window
Parameters:
<Mode> AC
AC coupling: Frequency offset and phase offset are automati-
cally corrected and DC signal contains a phase offset of ±π.
DC
DC coupling: Phase runs according to the existing frequency off-
set.
*RST: AC
Example: //Select AF coupling mode
CONF:TRAN:AF:COUP AC
Manual operation: See "Selecting the AF coupling mode" on page 139
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint:MODE <Mode>
This command selects the zero phase positioning mode.