Remote Control
R&S
®
FSPN
389User Manual 1179.4363.02 ─ 01
Example: //Select a custom step size
FREQ:CENT:STEP:LINK OFF
FREQ:CENT:STEP:LINK:FACT 20
Manual operation: See "Defining a frequency step size" on page 133
[SENSe:]FREQuency:SPAN <Frequency>
This command defines the frequency span.
Parameters:
<Frequency> <numeric value>
Default unit: Hz
Example: //Define a frequency span
FREQ:SPAN 500MHZ
Manual operation: See "Defining the frequency range of the measurement"
on page 133
11.6.9.2 Amplitude Configuration
Commands to configure the amplitude described elsewhere.
●
Attenuation mode: INPut<1|2>:ATTenuation:AUTO
●
Attenuation level: INPut<1|2>:ATTenuation
●
Input coupling: INPut<ip>:COUPling
INPut<1|2>:RLEVel........................................................................................................ 389
INPut<1|2>:RLEVel <Reflevel>
This command defines the reference level (expected signal level).
Note that the reference level is coupled to the attenuation.
Suffix:
<1|2>
.
1|2
irrelevant
Parameters:
<Reflevel> <numeric value>
*RST: 0
Default unit: dBm
Example: //Define reference level
INP:RLEV 10DBM
Manual operation: See "Defining a reference level" on page 137
Common Measurement Settings