EasyManua.ls Logo

R&S FSPN - Page 399

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
399User Manual 1179.4363.02 ─ 01
Example:
CALC:SETT:TIME:TRAC Trace1
CALCulate<n>:SETTling:TIME:FREQuency:LIMit:LOWer <Frequency>
This command defines the lower frequency of settling time measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> Default unit: Hz
Example:
CALC1:SETT:TIME:LOW:FREQ:LIM 100
CALCulate<n>:SETTling:TIME:FREQuency:LIMit:UPPer <Frequency>
This command defines the upper frequency of settling time measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> Default unit: Hz
Example:
CALC1:SETT:TIME:UPP:FREQ:LIM 1000
CALCulate<n>:SETTling:TIME:RESult?
This command returns the settling time.
Suffix:
<n>
.
irrelevant
Example:
CALC1:SETT:TIME:RES?
Usage: Query only
11.6.9.7 Frequency Deviation
CALCulate<n>:FREQuency:DEViation:STATe...................................................................399
CALCulate<n>:FREQuency:DEViation:DIAGram...............................................................400
CALCulate<n>:FREQuency:DEViation:TRACe..................................................................400
CALCulate<n>:FREQuency:DEViation:STARt...................................................................400
CALCulate<n>:FREQuency:DEViation:STOP....................................................................400
CALCulate<n>:FREQuency:DEViation:REGRession?........................................................401
CALCulate<n>:FREQuency:DEViation:REGRession:POINts?............................................ 401
CALCulate<n>:FREQuency:DEViation:SLOPe?................................................................401
CALCulate<n>:FREQuency:DEViation:STATe <State>
This command defines the state of frequency deviation measurement.
Common Measurement Settings

Table of Contents