Remote Control
R&S
®
FSPN
508User Manual 1179.4363.02 ─ 01
function group number
board number
function number
parameter 1 (see the Service Manual)
parameter 2 (see the Service Manual)
Example:
DIAG:SERV:SFUN 'Function1'
DIAG:SERV:SFUN? 'Function2'
Manual operation: See " Service Function " on page 240
See " Send " on page 241
DIAGnostic:SERVice:SFUNction:LASTresult?
This command queries the results of the most recent service function you have used.
Return values:
<Result>
Usage: Query only
DIAGnostic:SERVice:SFUNction:RESults:DELete
This command deletes the results of the most recent service function you have used.
Usage:
Event
Manual operation: See " Clear Results " on page 241
DIAGnostic:SERVice:SFUNction:RESults:SAVE [<FileName>]
This command saves the results of the most recent service function you have used.
Parameters:
<FileName> String containing the file name.
Manual operation: See " Save Results " on page 241
DIAGnostic:SERVice:SINFo?
This command creates a *.zip file with important support information. The *.zip file con-
tains the system configuration information ("device footprint"), the current eeprom data
and a screenshot of the screen display (if available).
As a result of this command, the created file name (including the drive and path) is
returned.
If you contact the Rohde&Schwarz support to get help for a certain problem, send this
file to the support in order to identify and solve the problem faster.
Return values:
<FileName>
Configuring the R&S FSPN