Remote Control
R&S
®
FSPN
340User Manual 1179.4363.02 ─ 01
Example: //Query maximum current
SOUR:CURR:AUX:LIM:HIGH?
returns
0.1
Usage: Query only
SOURce:CURRent:CONTrol<i>:LIMit:HIGH?
This command returns the maximum current of the V
tune
connector.
Suffix:
<i>
.
1..n
irrelevant
Return values:
<Current> <numeric value>
The return value is always 0.02 A.
Default unit: A
Example: //Query maximum current
SOUR:CURR:CONT:LIM:HIGH?
returns
0.02
Usage: Query only
SOURce:CURRent:POWer<i>:LIMit:HIGH <Current>
This command defines the maximum current of the V
supply
connector.
Prerequisites for this command
V
supply
is controlled in terms of voltage (SOURce:VOLTage:POWer<i>:LEVel:
MODE).
Suffix:
<i>
.
1..n
irrelevant
Parameters:
<Current> <numeric value>
Range: 0 to 2
*RST: 2
Default unit: A
Example: //Define a maximum output current of 1000 mA
SOUR:CURR:POW:LIM:HIGH 1
SOURce:CURRent:SEQuence:RESult?
This command queries the actually measured current on the DC power sources.