Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Turn on the V
supply
voltage source
SOUR:VOLT:POW:LEV ON
Manual operation: See "Turning the output on and off" on page 124
SOURce:VOLTage:POWer<i>:LIMit:HIGH <Voltage>
This command defines the maximum current or voltage that may be supplied by the
V
supply
source.
Suffix:
<i>
.
1..n
irrelevant
Parameters:
<Voltage> <numeric value>
The type of value depends on whether you control the output in
terms of current or voltage (SOURce:VOLTage:POWer<i>:
LEVel:MODE).
When you control it in terms of voltage, the value is a current
(A). When you control it in terms of current, the value is a volt-
age (V).
Range: 0 to 16 V or 2000 mA
*RST: 16 V or 2000 mA
Default unit: V or A
Example: //Define a maximum output of 1000 mA
SOUR:VOLT:POW:LEV:MODE VOLT
SOUR:VOLT:POW:LIM:HIGH 1000MA
Manual operation: See "Defining the output level" on page 125
SOURce:VOLTage:SEQuence:RESult?
This command queries the actually measured voltages on the DC power sources.
Prerequisites for this command
Turn on DC power sources (SOURce:VOLTage[:STATe]).
Example:
//Query voltages of the DC power sources
SOUR:VOLT:SEQ:RES?
would return, e.g.
0.000000, -10.000000, -10.000000
Usage: Query only
Manual operation: See "Reading the actual output levels" on page 126