SINGle
Displays the measurement channel that was previously focused.
*RST: SING
Example:
DISP:FORM SPL
DISPlay:WSELect?
This command queries the currently active window (the one that is focused) in the cur-
rently selected measurement channel.
Return values:
<SelectedWindow> Index number of the currently active window.
Range: 1 to 16
Usage: Query only
DISPlay[:WINDow<n>][:SUBWindow<w>]:SELect
This command sets the focus on the selected result display window.
This window is then the active window.
For measurements with multiple results in subwindows, the command also selects the
subwindow. Use this command to select the (sub)window before querying trace data.
Suffix:
<n>
.
Window
<w> subwindow
Not supported by all applications
Example: //Put the focus on window 1
DISP:WIND1:SEL
Example: //Put the focus on subwindow 2 in window 1
DISP:WIND1:SUBW2:SEL
DISPlay[:WINDow<n>]:SIZE <Size>
This command maximizes the size of the selected result display window temporarily.
To change the size of several windows on the screen permanently, use the LAY:SPL
command (see LAYout:SPLitter on page 406).
Suffix:
<n>
.
Window
Parameters:
<Size> LARGe
Maximizes the selected window to full screen.
Other windows are still active in the background.