When the instrument operates in remote lock mode, all the keys on the front
panel cannot be used except for the output on/off key of each channel , ,
and the power switch key . At this point, you can press the output on/off
key only to disable channel output.
If <bool> is omitted, the command enables the remote lock.
You can also send
:SYSTem:KLOCk:STATe
to enable or disable remote lock.
Return Format
The query returns 1 or 0.
Examples
:SYST:RWL ON /*Enables the remote lock.*/
4.11.16 :SYSTem:SAMPling
Syntax
:SYSTem:SAMPling <
mode
>
:SYSTem:SAMPling?
Description
Sets or queries the sampling mode.
Parameter
Name Type Range Default
<mode> Discrete {AUTO|HIGH|LOW} AUTO
Remarks
DP2000 series power supply provides three current sampling modes for CH1 and
CH2. You can select “High Curr, “Low Curr, or “AUTO”, and the default is
AUTO”.
- HIGH: The power supply uses the high range current mode.
- LOW: The maximum current that can be programmed is 0.5 A and the
maximum current that can be read is 11 mA. When 11 mA is reached, the
Command System