EasyManuals Logo

Rigol DP2000 Series Programming Guide

Rigol DP2000 Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Description
Waits for all the pending operations to complete before executing any other
commands.
Parameter
None.
Remarks
When "BUS" (Bus trigger, namely software trigger) is selected, this command can
ensure synchronization. After the command is executed, the instrument will wait for
all the pending operations to complete before executing any other commands.
Return Format
None.
Examples
*WAI /*Waits for all the pending operations to complete before
executing any other commands.*/
4.4 :INSTrument Commands
:INSTrument commands are used to select the channel to be programmed or query
the channel currently selected.
4.4.1 :INSTrument:NSELect
Syntax
:INSTrument:NSELect <
n
>
:INSTrument:NSELect?
Description
Selects the channel to be programmed or queries the channel currently selected.
Parameter
Name Type Range Default
<n> Discrete {1|2|3} 1
Remarks
The parameters 1, 2, and 3 represent CH1, CH2, and CH3 respectively.
This command uses numbers to substitute the channel identifiers
in
:INSTrument[:SELEct]
. It functions the same as
:INSTrument[:SELEct]
and
:INSTrument[:SELect]
.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
31

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals