4.10.9 :STATus:QUEStionable:INSTrument:ISUMmary[<n>]:COND
ition?
Syntax
:STATus:QUEStionable:INSTrument:ISUMmary[<
n
>]:CONDition?
Description
Sets or queries the output mode for the specified channel.
Parameter
Name
Type Range Default
<n> Integer {1|2|3} -
Remarks
If [<n>] is omitted, the command queries the output mode of the current
channel.
Execute the command and the query returns +0, +1, +2, or +3, as shown in the
table below.
Returned Value Description
+0 The output is off.
+1 The output is in CC (constant current) mode.
+2 The output is in CV (constant voltage) mode.
+3 The output is in UR (unregulated) mode.
Return Format
The query returns +0, +1, +2, or +3.
Examples
:STAT:QUES:INST:ISUM1:COND? /*Queries the output mode of CH1. The
query returns +1.*/
4.10.10 :STATus:QUEStionable:INSTrument:ISUMmary[<n>]:ENABl
e
Syntax
:STATus:QUEStionable:INSTrument:ISUMmary[<
n
>]:ENABle <
enable value
>
:STATus:QUEStionable:INSTrument:ISUMmary[<
n
>]:ENABle?
Command System
74
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals