Return Format
The query returns 1 or 2.
Examples
:SYST:COMM:RS232:SBIT 2 /*Sets the stop bit to 2.*/
:SYST:COMM:RS232:SBIT? /*Queries the stop bit. The query returns
2.*/
4.11.8 :SYSTem:ERRor[:NEXT]?
Syntax
:SYSTem:ERRor[:NEXT]?
Description
Queries and removes errors from the error queue.
Parameter
None.
Remarks
The power supply beeps once each time an error is detected. If more than 20 errors
have occurred, the last error stored in the queue (the most recent error) is replaced
with -350, “Queue overflow”. No additional errors are stored until you remove
errors from the queue.
Errors are retrieved in first-in-first-out (FIFO) order.
The error queue is cleared when power has been off or after
*CLS
has been executed.
The
*RST
command does not clear the error queue.
Return Format
The query returns the name and content of the error message, for example,
-113,"Undefined header; keyword cannot be found”. If no error has occurred, the
query returns 0,“No error.
Examples
None.
4.11.9 :SYSTem:KLOCk:STATe
Syntax
:SYSTem:KLOCk:STATe <
bool
>
:SYSTem:KLOCk:STATe?
Command System
90
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals