EasyManuals Logo

Rigol DP2000 Series Programming Guide

Rigol DP2000 Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
Description
Queries whether an overvoltage protection (OVP) event occurred on the specified
channel.
Parameter
Name Type Range Default
<n> Integer {1|2|3} -
Remarks
The overvoltage protection (OVP) function disables the output automatically
when the actual output voltage reaches the OVP level.
If [:SOURce[<n>]] or [<n>] is omitted, the command queries whether an OVP
event occurred on the current channel.
You can also send
:OUTPut:OVP:ALAR?
or
:OUTPut:OVP:QUES?
to query whether
an OVP event has occurred on the specified channel.
You can send
[:SOURce[<n>]]:VOLTage:PROTection:CLEar
to clear an OVP event
that occurred on the specified channel.
Return Format
The query returns 1 or 0.
Examples
:VOLT:PROT:TRIP? /*Queries whether an OVP event occurred on the
current channel. The query returns 1.*/
4.10 :STATus Commands
:STATus commands are used to set and query the Questionable Status register and
Operation Status register.
4.10.1 :STATus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
Queries the value of the condition register of the Operation Status register.
Parameter
None.
Command System
68
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals