Description
Queries whether an overvoltage protection (OVP) event occurred on the specified
channel.
Parameter
Name Type Range Default
<n> Integer {1|2|3} -
Remarks
• The overvoltage protection (OVP) function disables the output automatically
when the actual output voltage reaches the OVP level.
• If [:SOURce[<n>]] or [<n>] is omitted, the command queries whether an OVP
event occurred on the current channel.
• You can also send
:OUTPut:OVP:ALAR?
or
:OUTPut:OVP:QUES?
to query whether
an OVP event has occurred on the specified channel.
• You can send
[:SOURce[<n>]]:VOLTage:PROTection:CLEar
to clear an OVP event
that occurred on the specified channel.
Return Format
The query returns 1 or 0.
Examples
:VOLT:PROT:TRIP? /*Queries whether an OVP event occurred on the
current channel. The query returns 1.*/
4.10 :STATus Commands
:STATus commands are used to set and query the Questionable Status register and
Operation Status register.
4.10.1 :STATus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
Queries the value of the condition register of the Operation Status register.
Parameter
None.
Command System
68
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.