Parameter
Name Type Range Default
<source> Discrete {CH1|CH2|CH3} -
<bool> Bool {0|1|ON|OFF} 0
Remarks
• When OVP is enabled, the output is disabled automatically if the output voltage
reaches the OVP level (
:OUTPut:OVP:VALue
) currently set. You can
send
:OUTPut:OVP:QUES?
or
:OUTPut:OVP:ALAR?
to query whether an OVP
event occurred on the specified channel.
• If <source> is omitted, the command executes the corresponding operation on
the current channel.
Return Format
The query returns 1 or 0.
Examples
:OUTP:OVP CH1,1 /*Enables the OVP function for CH1.*/
:OUTP:OVP? CH1 /*Queries the OVP on/off state for CH1. The query
returns 1.*/
4.8.13 :OUTPut:OVP:VALue
Syntax
:OUTPut:OVP:VALue [<
source
>,]{<
value
>|<
lim
>}
:OUTPut:OVP:VALue? [<
source
>][,<
lim
>]
:OUTPut:OVP:VALue? [<
lim
>]
Description
Sets or queries the overcurrent protection (OVP) level for the specified channel.
Parameter
Name Type Range Default
<source> Discrete {CH1|CH2|CH3} -
<value> Real
Please refer to
Table 4.33 Range and default value
of overvoltage/overcurrent protection
<lim> Discrete {MINimum|MAXimum} -
Command System
54