EasyManua.ls Logo

Rigol DP2000 Series - Timer:templet:points

Rigol DP2000 Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter
Name Type Range Default
<value> Real 0.001 s to 3600 s
Refer to
Remarks
Remarks
<value> specifies the duration of a cycle for a waveform. All waveforms except
for ExpRise and ExpFall require a period.
By default, the period of Pulse is 2 s, and the period of other waveforms is 50 s.
The total number of points within a period of the selected template is
determined by the period of the waveform and time interval
(
:TIMEr:TEMPlet:INTErval
). The number of points within a period=period/time
interval.
Return Format
The query returns a real number ranging from 0.001 to 3600, for example, 15.000.
Examples
:TIME:TEMP:PERI 15 /*Sets the Period to 15 s.*/
:TIME:TEMP:PERI? /*Queries the Period for the selected template.
The query returns 15.000.*/
4.12.18 :TIMEr:TEMPlet:POINTs
Syntax
:TIMEr:TEMPlet:POINTs <
value
>
:TIMEr:TEMPlet:POINTs?
Description
Sets or queries the total number of points.
Parameter
Name Type Range Default
<value> Integer 1 to 512 50
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
113

Table of Contents

Other manuals for Rigol DP2000 Series

Related product manuals