Parameter
Name Type Range Default
<n> Integer {1|2|3} -
<numeric value> Real
0 to the maximum current
value of the specified channel
Refer to
Remarks
<def> Discrete {DEFault} -
Remarks
If [:SOURce[<n>]] or [<n>] is omitted, the command sets the corresponding
parameter of the channel currently selected.
<numeric value> is the step size specified. DEFault is the default value. The
default values of <numeric value> are as shown in the table below.
Channel Default
CH1 0.0001 A
CH2 0.0001 A
CH3 0.001 A
Select UP or DOWN and executes
[:SOURce[<n>]]:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
, and the instrument will step up or down the current by the step
size set in this command.
Return Format
The query returns the step size, for example, 0.1000.
Examples
:CURR:STEP 0.1 /*Sets the step size to 0.1 A for the selected
channel.*/
:CURR:STEP? /*Queries the step size for the selected channel. The
query returns 0.1000.*/
4.9.3 [:SOURce[<n>]]:CURRent:PROTection:CLEar
Syntax
[:SOURce[<
n
>]]:CURRent:PROTection:CLEar
Description
Clears an OCP event that occurred on the specified channel and enables the output of
the corresponding channel.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
59

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals