Examples
:TIMEr:CHANNel CH2 /*Sets the channel currently edited to CH2.*/
:TIMEr:CHANNel? /*Queries the channel currently edited. The query
returns CH2.*/
4.12.3 :TIMEr:ENDState
Syntax
:TIMEr:ENDState <
end
>
:TIMEr:ENDState?
Description
Sets or queries the end state of the generator.
Parameter
Name Type Range Default
<end> Discrete {OFF|LAST} OFF
Remarks
"End state" refers to the state of the instrument after it has completed outputting
groups of voltage/current when the number of cycles is finite.
• OFF: the instrument shuts down output automatically once output is completed.
• LAST: the instrument stays in the output state of the last group after the output
is completed.
The total number of groups in timing output = the number of output groups × the
number of cycles; wherein, you can send
:TIMEr:GROUPs:NUM?
to query the number
of output groups and send
:TIMEr:CYCLEs
to set the number of cycles.
Return Format
The query returns OFF or LAST.
Examples
:TIME:ENDS LAST /*Sets the end state to “LAST”.*/
:TIME:ENDS? /*Queries the end state. The query returns LAST.*/
4.12.4 :TIMEr:GROUPs:NUM?
Syntax
:TIMEr:GROUPs:NUM?
Description
Queries the number of output groups in the Arb editor.
Command System