Queries the current operation mode.
Parameter
Name Type Range Default
<mode> Discrete {LOCal|REMote|RWLock} LOCal
Remarks
• LOCal: Local mode.
• REMote: Remote mode. The touch screen and all front-panel keys are disabled
except for the front-panel output on/off keys and key. At this point, you can
press the output on/off key only to disable channel output.
• RWLock: Remote lock mode. The touch screen and all front-panel keys are
disabled except for the front-panel output on/off keys. At this point, you can
press the output on/off key only to disable channel output. You can only use the
specified command to disable the remote lock mode.
Return Format
The query returns LOCal, REMote, or RWLock.
Examples
:SYSTem:COMMunicate:RLSTate REMote /*Sets the power supply to
remote mode.*/
:SYSTem:COMMunicate:RLSTate? /*Queries the operation mode. The
query returns REMote.*/
4.11.7 :SYSTem:COMMunicate:RS232
4.11.7.1 :SYSTem:COMMunicate:RS232:BAUD
Syntax
:SYSTem:COMMunicate:RS232:BAUD <
baud
>
:SYSTem:COMMunicate:RS232:BAUD?
Description
Sets or queries the baud rate of the RS232 interface. The unit is Baud.
Command System