4.3.9 *RST
Syntax
*RST
Description
Restores the power supply to its factory default.
Parameter
None.
Remarks
Executing this command will immediately restore the power supply to its factory
default without querying.
Return Format
None.
4.3.10 *SAV
Syntax
*SAV <
n
>
Description
Saves the current instrument state to the specified location in the internal memory
with the specified filename (RIGOLn.RSF).
Parameter
Name
Type Range Default
<n> Discrete {0|1|2|3|4|5|6|7|8|9} -
Remarks
The command saves the current instrument state in the specified location,
overwriting the previous state of the same filename (if any). If the state file
stored in the specified storage location is locked (
:MEMory:LOCK
), this command
is invalid (not overwrite the previous file).
You can send
:MEMory:STORe
to save the current status to internal/external
memory.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.