Parameter
Name Type Range Default
<bool> Bool {0|1|ON|OFF} 0|OFF
Remarks
• Turning on the generator will change the channel output state. Please make sure
that the change in the output state will not affect the device connected to the
power supply before enabling the output.
• The arbitrary waveform output is valid only when both the generator and the
selected channel are turned on.
• When the run mode (
:TIMEr:RUN
) is set to “Continuous” (CONTinue), turning
on the selected channel and the generator (:TIME ON), the instrument will repeat
the sequence continuously based on the number of data groups and repetition
cycle currently set. If the trigger source (
:TIMEr:TRIG
) is set to “BUS”, you also
need to send
*TRG
to enable the output.
• When the run mode (
:TIMEr:RUN
) is set to “Single” (SINGle), turning on the
selected channel and the generator (:TIME ON), the instrument will output a
single group of data in order each time :TIME ON is sent. If the trigger source
(
:TIMEr:TRIG
) is set to “BUS”, the instrument will output a single group of data
each time
*TRG
is sent.
• While the generator is turned on, Arb parameters cannot be modified.
Return Format
The query returns 1 or 0.
Examples
:TIME ON /*Turns on the generator.*/
:TIME? /*Queries the run/stop state of the generator. The query
returns 1.*/
4.12.10 :TIMEr:TEMPlet:CONSTruct
Syntax
:TIMEr:TEMPlet:CONSTruct
Description
Sets the Arb parameters based on the template currently selected and the parameters
set.
Parameter
None.
Command System