• Select UP or DOWN and executes
[:SOURce[<n>]]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude]
, and the instrument will step up or down the voltage by the step
size set in this command.
Return Format
The query returns the step size, for example, 0.100.
Examples
:VOLT:STEP 0.1 /*Sets the step size of voltage to 0.1 V for the
current channel.*/
:VOLT:STEP? /*Queries the step size of voltage for the current
channel. The query returns 0.100.*/
4.9.9 [:SOURce[<n>]]:VOLTage:PROTection:CLEar
Syntax
[:SOURce[<
n
>]]:VOLTage:PROTection:CLEar
Description
Clears an OVP event that occurred on the specified channel and enables the output of
the corresponding channel.
Parameter
Name
Type Range Default
<n> Integer {1|2|3} -
Remarks
• You can send
[:SOURce[<n>]]:VOLTage:PROTection:TRIPped?
to query whether
an OVP event occurred on the specified channel.
• Before executing the command, make sure to remove the condition that caused
the overvoltage protection on the specified channel (you can either decrease the
output voltage to below the OVP level or increase the OVP level to be greater
than the output voltage). Execute this command to clear an OVP event that
occurred on the specified channel and enable the output of the corresponding
channel.
• If [:SOURce[<n>]] or [<n>] is omitted, the command clears an OVP event
occurred on the current channel.
• You can send
:OUTPut:OVP:CLEar
command to clear an OVP event that occurred
on the specified channel.
Return Format
None.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
65