output current is displayed as “-.----” in the interface. When the power
supply outputs current within 11 mA, it uses the low range current mode.
- AUTO: When the power supply outputs current higher than or equal to 11
mA, it uses the high range current mode; when the power supply outputs
current lower than 11 mA, it automatically switches to the low range current
measurement mode.
The three sampling modes are only available for CH1 and CH2. CH3 has no low
current sampling mode, and hence cannot switch sampling mode.
When the instrument operates in parallel mode (internal), the sampling mode is
fixed to high range current mode. If you want to use external parallel connection,
you need to set the sampling mode to high current mode manually.
Return Format
The query returns AUTO, HIGH, or LOW.
Examples
:SYSTem:SAMPling HIGH /*Sets the sampling mode to high range mode.*/
:SYSTem:SAMPling? /*Queries the sampling mode. The query returns
HIGH.*/
4.11.17 :SYSTem:SAVer
Syntax
:SYSTem:SAVer <
bool
>
:SYSTem:SAVer?
Description
Sets or queries the on/off state of the screen saver function.
Parameter
Name Type Range Default
<bool> Bool {1|0|ON|OFF} 1
Remarks
None.
Command System
96
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals