Page #117 background imageLoading...
Page #117 background image
Description
Sets or queries the Time Interval.
Parameter
Name Type Range Default
<value> Real 0.001 s to 3600 s 1 s
Remarks
Time interval refers to the time required for the instrument to output each group
of parameters using the template currently selected.
This command is not available for Stair Up, Stair Dn, Stair UpDn, and Pulse.
Return Format
The query returns a real number, for example, 15.000.
Examples
:TIME:TEMP:INTE 15 /*Sets the Time Interval to 15 s.*/
:TIME:TEMP:INTE? /*Queries the Time Interval currently set. The
query returns 15.000.*/
4.12.13 :TIMEr:TEMPlet:INVErt
Syntax
:TIMEr:TEMPlet:INVErt <
bool
>
:TIMEr:TEMPlet:INVErt?
Description
Sets or queries the on/off state of the invert function of the template currently
selected.
Parameter
Name Type Range Default
<bool> Bool {0|1|ON|OFF} OFF
Remarks
When the invert function is enabled, the instrument will first invert the waveform
and then set the waveform parameters.
The invert function is available for Sine, Pulse, and Ramp templates only.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
109

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals