EasyManuals Logo

Rigol DP2000 Series Programming Guide

Rigol DP2000 Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
Parameter
Name Type Range Default
<d> Discrete {D1|D2|D3|D4} D1
<bool> Bool {1|ON|0|OFF} 0|OFF
Remarks
When the specified data line receives input signal that meets the current trigger type
(
:TRIGger:IN:TYPE
), the specified source under control (
:TRIGger:IN:SOURce
) will turn
on/off the output, or toggle the output state according to the setting
in
:TRIGger:IN:RESPonse
.
Return Format
The query returns 1 or 0.
Examples
:TRIG:IN D1,ON /*Enables the trigger input function for D1.*/
:TRIG:IN? D1 /*Queries the on/off state of the trigger input
function for D1. The query returns 1.*/
4.13.2 :TRIGger:IN:IMMEdiate
Syntax
:TRIGger:IN:IMMEdiate
Description
As soon as the trigger system is initiated, the analog hardware will send the trigger
signal immediately.
Parameter
None.
Remarks
None.
Return Format
None.
Examples
None.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
119

Table of Contents

Other manuals for Rigol DP2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP2000 Series and is the answer not in the manual?

Rigol DP2000 Series Specifications

General IconGeneral
BrandRigol
ModelDP2000 Series
CategoryPower Supply
LanguageEnglish

Related product manuals