RIGOL Chapter 2 Command System
2-122 DSA800E Programming Guide
[:SENSe]:DETector[:FUNCtion]
Syntax
[:SENSe]:DETector[:FUNCtion] NEGative|NORMal|POSitive|RMS|SAMPle|VAVerage|QPEak
[:SENSe]:DETector[:FUNCtion]?
Description
Set the detector type.
Query the detector type.
Parameter
-- Keyword
NEGative|NORMal|POSitive|
RMS|SAMPle|VAVerage|QPEak
POSitive
Explanation
NEGative: negative peak
NORMal: normal detector
POSitive: positive peak
RMS: RMS average
SAMPle: sample detector
VAVerage: voltage average
QPEak: Quasi-Peak
Return Format
The query returns NEG, NORM, POS, RMS, SAMP, VAV or QPEAK.
Example
The command below sets the detector type to Quasi-Peak.
:SENSe:DETector:FUNCtion QPEak
The query below returns QPEAK.
:SENSe:DETector:FUNCtion?
[:SENSe]:EBWidth:AVERage:COUNt
Syntax
[:SENSe]:EBWidth:AVERage:COUNt <integer>
[:SENSe]:EBWidth:AVERage:COUNt?
Description
Set the number of averages of the emission bandwidth measurement.
Query the number of averages of the emission bandwidth measurement.
Parameter
Explanation
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query returns the number of averages in integer.