RIGOL Chapter 2 Command System
2-190 DSA800E Programming Guide
:TRIGger:SEQuence:SOURce
Syntax
:TRIGger:SEQuence:SOURce IMMediate|VIDeo|EXTernal
:TRIGger:SEQuence:SOURce?
Description
Set the trigger type.
Query the trigger type.
Parameter
Name
Type
Range
Default
--
Keyword
IMMediate|VIDeo|EXTernal
IMMediate
Explanation
IMMediate: free run trigger
VIDeo: video trigger. This trigger type is not available in non-zero span mode as well as RMS Avg detection
or Voltage Avg detection in zero span mode.
EXTernal: external trigger
Return Format
The query returns IMM, VID or EXT.
Example
The command below sets the trigger type to external trigger.
:TRIGger:SEQuence:SOURce EXTernal
The query below returns EXT.
:TRIGger:SEQuence:SOURce?