RIGOL Chapter 2 Command System
2-134 DSA800E Programming Guide
[:SENSe]:HDISt:TIME
Syntax
[:SENSe]:HDISt:TIME <time>
[:SENSe]:HDISt:TIME?
Description
Set the sweep time of the harmonic distortion measurement.
Query the sweep time of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
20 us to 3.2 ks
16 ms
Explanation
This command is only available when the harmonic distortion measurement is enabled.
Return Format
The query returns the sweep time in scientific notation and the unit is s.
Example
The command below sets the sweep time to 100 ms.
:SENSe:HDISt:TIME 0.1
The query below returns 1.000000E-01.
:SENSe:HDISt:TIME?
[:SENSe]:HDISt:TIME:AUTO[:STATe]
Syntax
[:SENSe]:HDISt:TIME:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:HDISt:TIME:AUTO[:STATe]?
Description
Enable or disable the auto setting mode of the sweep time of the harmonic distortion measurement.
Query the status of the auto setting mode of the sweep time of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Explanation
This command is only available when the harmonic distortion measurement is enabled.
This setting changes the sweep time of the spectrum analyzer.
Return Format
The query returns 0 or 1.
Example
The command below enables the auto setting mode of the sweep time.
:SENSe:HDISt:TIME:AUTO:STATe ON or :SENSe:HDISt:TIME:AUTO:STATe 1
The query below returns 1.
:SENSe:HDISt:TIME:AUTO:STATe?