Chapter 2 Command System RIGOL
DSA800E Programming Guide 2-139
[:SENSe]:POWer:ATUNe
Syntax
[:SENSe]:POWer:ATUNe
Description
Search for signals within the full frequency range and adjust the frequency and amplitude for optimum
display effect of the signal.
[:SENSe]:POWer[:RF]:ATTenuation
Syntax
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Description
Set the attenuation of the RF attenuator.
Query the attenuation of the RF attenuator.
Parameter
Name
Type
Range
Default
<rel_ampl>
Integer
0 dB to 30 dB
10 dB
Return Format
The query returns the attenuation in integer and the unit is dB.
Example
The command below sets the attenuation to 20 dB.
:SENSe:POWer:RF:ATTenuation 20
The query below returns 20.
:SENSe:POWer:RF:ATTenuation?

Table of Contents