RIGOL Chapter 2 Command System
2-150 DSA800E Programming Guide
[:SENSe]:TPOWer:RLIMit
Syntax
[:SENSe]:TPOWer:RLIMit <time>
[:SENSe]:TPOWer:RLIMit?
Description
Set the stop line of the T-power measurement.
Query the stop line of the T-power measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
Start line to the sweep time
50 ms
Explanation
This command is only available when the T-power measurement is enabled.
Return Format
The query returns the stop line in scientific notation and the unit is s.
Example
The command below sets the stop line to 10 ms.
:SENSe:TPOWer:RLIMit 0.01
The query below returns 1.000000E-02.
:SENSe:TPOWer:RLIMit?
[:SENSe]:VSWR:FREFlect
Syntax
[:SENSe]:VSWR:FREFlect
Description
The spectrum analyzer executes the first measurement (with the device under test disconnected) of the
VSWR measurement.
Explanation
This command is only available when the VSWR measurement is enabled.
Disconnect the device under test before sending this command.
[:SENSe]:VSWR:NREFlect
Syntax
[:SENSe]:VSWR: NREFlect
Description
The spectrum analyzer executes the second measurement of the VSWR measurement.
Explanation
This command is only available when the VSWR measurement is enabled.
Disconnect the device under test and send the [:SENSe]:VSWR:FREFlect command; then connect the the
device under test and send this command.

Table of Contents

Other manuals for Rigol DSA832E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832E and is the answer not in the manual?

Rigol DSA832E Specifications

General IconGeneral
BrandRigol
ModelDSA832E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals