Chapter 2 Command System RIGOL
DSA800E Programming Guide 2-123
Example
The command below sets the number of averages to 100.
:SENSe:EBWidth:AVERage:COUNt 100
The query below returns 100.
:SENSe:EBWidth:AVERage:COUNt?
[:SENSe]:EBWidth:AVERage[:STATe]
Syntax
[:SENSe]:EBWidth:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:EBWidth:AVERage[:STATe]?
Description
Enable or disable the average measurement function of the emission bandwidth measurement.
Query the status of the average measurement function of the emission bandwidth measurement.
Parameter
Explanation
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query returns 0 or 1.
Example
The command below enables the average measurement function.
:SENSe:EBWidth:AVERage:STATe ON or :SENSe:EBWidth:AVERage:STATe 1
The query below returns 1.
:SENSe:EBWidth:AVERage:STATe?
[:SENSe]:EBWidth:AVERage:TCONtrol
Syntax
[:SENSe]:EBWidth:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:EBWidth:AVERage:TCONtrol?
Description
Set the average mode of the emission bandwidth measurement.
Query the average mode of the emission bandwidth measurement.
Parameter
Explanation
EXPonential: exponential average
REPeat: repeat average
When exponential average is selected, the result is the exponential average of the latest N (specified by the