RIGOL Chapter 2 Command System
2-12 DSA800E Programming Guide
:CALCulate:MARKer:FCOunt:RESolution:AUTO
Syntax
:CALCulate:MARKer:FCOunt:RESolution:AUTO OFF|ON|0|1
:CALCulate:MARKer:FCOunt:RESolution:AUTO?
Description
Set the resolution of the frequency counter automatically.
Query the status of the auto setting of the resolution of the frequency counter.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The command below disables the auto setting of the resolution of the frequency counter.
:CALCulate:MARKer:FCOunt:RESolution:AUTO OFF or :CALCulate:MARKer:FCOunt:RESolution:AUTO 0
The query below returns 0.
:CALCulate:MARKer:FCOunt:RESolution:AUTO?
:CALCulate:MARKer:FCOunt:X?
Syntax
:CALCulate:MARKer:FCOunt:X?
Description
Query the reading of the frequency counter.
Return Format
The query returns the reading in integer and the unit is Hz.
The query returns 9000000000000000 when the frequency counter is disabled.