:CALibration Subsystem
Command List:
:CALibration:[ALL]
:CALibration:AUTO
:CALibration:[ALL]
Syntax
:CALibration:[ALL]
Description
Execute self-calibration immediately.
Example
The command below makes the instrument execute self-calibration immediately.
:CALibration:ALL
:CALibration:AUTO
Syntax
:CALibration:AUTO OFF|ON|0|1
:CALibration:AUTO?
Description
Enable or disable auto calibration.
Query the status of auto calibration.
Parameter
Explanation
By default, auto calibration is enabled. But the instrument will use the last setting at the next power-on if
users change this setting.
Return Format
The query returns 0 or 1.
Example
The command below enables auto calibration.
:CALibration:AUTO ON or :CALibration:AUTO 1
The query below returns 1.
:CALibration:AUTO?