EasyManua.ls Logo

Rigol DSA832E

Rigol DSA832E
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Command System
2-192 DSA800E Programming Guide
:UNIT Subsystem
Command List:
:UNIT:POWer
:UNIT:POWer
Syntax
:UNIT:POWer DBM|DBMV|DBUV|V|W
:UNIT:POWer?
Description
Set the unit of the Y axis.
Query the unit of the Y axis.
Parameter
Name
Type
Range
Default
--
Keyword
DBM|DBMV|DBUV|V|W
Refer to Explanation
Explanation
The default log unit is dBm.
The default linear unit is V.
Return Format
The query returns DBM, DBMV, DBUV, V or W.
Example
The comamnd below sets the amplitude unit to Watts.
:UNIT:POWer W
The query below returns W.
:UNIT:POWer?

Table of Contents

Other manuals for Rigol DSA832E

Related product manuals