RIGOL Chapter 2 Command System
2-192 DSA800E Programming Guide
:UNIT Subsystem
Command List:
:UNIT:POWer
:UNIT:POWer
Syntax
:UNIT:POWer DBM|DBMV|DBUV|V|W
:UNIT:POWer?
Description
Set the unit of the Y axis.
Query the unit of the Y axis.
Parameter
Name
Type
Range
Default
--
Keyword
DBM|DBMV|DBUV|V|W
Refer to Explanation
Explanation
The default log unit is dBm.
The default linear unit is V.
Return Format
The query returns DBM, DBMV, DBUV, V or W.
Example
The comamnd below sets the amplitude unit to Watts.
:UNIT:POWer W
The query below returns W.
:UNIT:POWer?