EasyManua.ls Logo

Rigol DSA832E

Rigol DSA832E
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Command System RIGOL
DSA800E Programming Guide 2-127
[:SENSe]:FREQuency:CENTer:DOWN
Syntax
[:SENSe]:FREQuency:CENTer:DOWN
Description
Decrease the center frequency at the center frequency step.
[:SENSe]:FREQuency:CENTer:SET:STEP
Syntax
[:SENSe]:FREQuency:CENTer:SET:STEP
Description
Set the CF step to the current center frequency.
[:SENSe]:FREQuency:CENTer:STEP:AUTO
Syntax
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Description
Enable or disable the auto setting mode of the CF step.
Query the status of the auto setting mode of the CF step.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The command below enables the auto setting mode.
:SENSe:FREQuency:CENTer:STEP:AUTO ON or :SENSe:FREQuency:CENTer:STEP:AUTO 1
The query below returns 1.
:SENSe:FREQuency:CENTer:STEP:AUTO?

Table of Contents