RIGOL Chapter 2 Command System
2-172 DSA800E Programming Guide
:SYSTem:OPTions?
Syntax
:SYSTem:OPTions?
Description
Query the option status (include the option number, order number, type and activation status) of the
spectrum analyzer.
Return Format
The query returns the option status (include the option number, order number, type and activation status)
of the spectrum analyzer. For example,
Serial NO.|Option|Option Type|Active
1| AMK-DSA800|Official|Y
2| EMI-DSA800|None|N
3| VSWR-DSA800|Official|Y
4| PA-DSA832|Official|Y
:SYSTem:PON:TYPE
Syntax
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?
Description
Set the instrument to recall the last setting or the preset setting at power-on.
Query the setting recalled by the instrument at power-on.
Parameter
Explanation
PRESet: preset setting, including factory setting and six user settings. Use the :SYSTem:PRESet:TYPE
command to select the desired setting.
LAST: the last setting.
Return Format
The query returns PRES or LAST.
Example
The command below sets the instrument to recall the last setting.
:SYSTem:PON:TYPE LAST
The query below returns LAST.
:SYSTem:PON:TYPE?