Chapter 2 Command System RIGOL
DSA800E Programming Guide 2-5
:CALCulate:LLINe:CONTrol:DOMain
Syntax
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Description
Set the X axis to denote frequency or time in the Pass/Fail test.
Query the type of the X axis in the Pass/Fail test.
Parameter
Name
Type
Range
Default
--
Keyword
FREQuency|TIME
FREQuency
Explanation
This setting is applicable to both the upper and lower limit lines.
All the points of the current limit line will be deleted when the X axis type is changed.
Return Format
The query returns FREQ or TIME.
Example
The command below sets the X axis to time.
:CALCulate:LLINe:CONTrol:DOMain TIME
The query below returns TIME.
:CALCulate:LLINe:CONTrol:DOMain?
:CALCulate:LLINe:FAIL?
Syntax
:CALCulate:LLINe:FAIL?
Description
Query the result of the Pass/Fail test.
Return Format
The query returns PASS or FAIL. The query returns UNMEAS if the test is not finished.